Made a new standard curve dilution. Ran just the curve to check it and compare to previous plates’ curves.
Serial Dilution of Curve info:
See previous post: FHL 2026 - qPCR prep and current status
Plate and Run info
Plate map: here
Ran 3.00+06 down to 3.00+00 in the standard curve (wells A1-A3, down to G1-G3).
I think in my previous dilution, I just didn’t go all the way down to 3.00+00 and only ran 3.00+07 down to 3.00+01 as my standard curve in wells A1-A3 down to G1-G3. This guess is based on the fact that when I went to make more dilutions, the last tube in the box was 3.00+08, when it should have been 3.00+07.
Results:
I set the threshold to 100, as that’s what I’ve been doing for all qPCR results.
I had bad replication for two of my dilutions, the initial results are the first image, then the filtered results are the second image:
Figure 1. Standard curve results on qPCR. Ran standards 3.00+06 down to 3.00+00. Set the threshold to 100.
Figure 2. Standard curve results on qPCR. Ran standards 3.00+06 down to 3.00+00. Set the threshold to 100. I removed the replicates that were way off, and the curves and run stats look great. But, moving forward, I need to make sure all triplicates look good.
I also did a comparison of the Cq values from this plate to a previous plate that used the previous dilution of the standard curve… and it appears that my assumptions are correct: I was using 3.00+07, 3.00+06, 3.00+05, 3.00+04, 3.00+03, 3.00+02, and 3.00+01 for my standard curve.
Figure 3. Cq values of the standard curve for plate 9 (previous standard curve that was used for all previous qPCR runs) and plate 10 (the new serial dilution of the standard curve started from 3.00+08 from the previous standard curve). This is showing that in all previous plates, my standard curve was actually 3.00+07 down to 3.00+01, not 3.00+06 down to 3.00+00.
Questions:
Moving forward, can I then just use 3.00+07–> 3.00+01 for my standard curve to keep it all consistent?
Would I have to re-adjust all previous plates to say that the standard curve was actually 3.00+07–> 3.00+01?